Asia Pacific Academy of Science Pte. Ltd. (APACSCI) specializes in international journal publishing. APACSCI adopts the open access publishing model and provides an important communication bridge for academic groups whose interest fields include engineering, technology, medicine, computer, mathematics, agriculture and forestry, and environment.

Modular implant design affects metal ion release following metal-on-metal hip arthroplasty: a retrospective study on 75 cases
Vol 33, Issue 2S1, 2019
Abstract
Metal-on-Metal (MoM) total hip arthroplasty (THA) has been associated to wear and metal-ions release, controversially related to a variety of clinical complications. Little is known about the relevant design-dependent parameters involved in this process. The present study investigated the correlation between metal ion release in blood and revision rate as a function of: (i) specific MoM implant modular design parameters, (i.e. acetabular cup and femoral head diameters, taper adapter material and size, femoral neck material and modularity and stem size); (ii) MoM bilaterality. Co and Cr ions concentration levels in blood of 75 patients were retrospectively-evaluated with a mean follow-up of 4.8 years (range: 1.8-6.3). Patients were divided in a unilateral and a bilateral group. Statistical analysis was performed to find any significant difference related to acetabular cup diameter, femoral head diameter, taper adapter material/size, neck material/size and stem size. The bilateral MoM group had 4-times higher metal ion levels in blood than the unilateral one (p=0.017 only Cr), related to a higher revision rate (30% vs 20%): differences were 10-times higher particularly with a 48 mm femoral head diameter (p=0.012) and a Ti-alloy neck (p=0.041). Within the monolateral group using a shorter taper adapter and a shorter neutrally-oriented neck demonstrated higher ion levels (p=0.038 only Cr and p=0.008 only Co, respectively). The aforementioned design-features and MoM bilaterality are important risk-factors for metal-ion release in modular MoM THA.
Keywords
References
Supporting Agencies
Copyright (c) 2019
This site is licensed under a Creative Commons Attribution 4.0 International License (CC BY 4.0).

Medical Genetics, University of Torino Medical School, Italy

Department of Biomedical, Surgical and Dental Sciences, University of Milan, Italy